於水清,赵珺彦.评价Canon TX-20非接触式眼压计准确性和可重复性[J].浙江中西医结合杂志,2015,25(8):
评价Canon TX-20非接触式眼压计准确性和可重复性
Accuracy and repeatability and of TX-20 tonometerYu Shuiqing, Zhao Junyan The First People"s Hospital of Xiaoshan District of Hangzhou, Hangzhou(311200), China
投稿时间:2014-11-08  修订日期:2014-12-18
DOI:
中文关键词:  Canon 非接触眼压计TX-20  Goldmann压平眼压计  准确性  可重复性
英文关键词:TX-20 non-contact tonometer  Goldmann applanation tonometer  Accuracy  Repeatability
基金项目:
作者单位E-mail
於水清* 杭州市萧山区第一人民医院 7393054@qq.com 
赵珺彦   
摘要点击次数: 1153
全文下载次数: 0
中文摘要:
      目的 对比Canon TX-20非接触式眼压计(TX-20)与Goldmann压平式眼压计(GAT)测量眼压的值,评价TX-20测量眼压的准确性和可重复性。方法 录选眼科就诊的60例患者共120只眼,其中男30例,女30例,年龄18岁~76岁。由同一检查者分别用两种眼压计连续测量3次。以GAT眼压值作为基准分为A、B、C 3组:A组(低眼压组)IOP<11 mmHg(1 kPa=7. 5 mmHg),B组(正常眼压组)IOP为11~21 mmHg,C组(高眼压组)IOP>21 mmHg。以Goldmann眼压计为标准,评价TX-20眼压计的准确性。分组分析TX-20连续3次测量值之间的可重复性。结果 与GAT测量值比较,TX-20A、B、C3组测量眼压值均偏低,二者差值95%confidence interval(95%CI)分别是0.31~2.95mmHg,1.09~4.38mmHg,0.28~7.28mmHg。A、B、C3组测量眼压的可重复性较好,3次测量结果之间差异均无统计学意义(P均>0.05)。结论 在低眼压和正常眼压组,TX-20测量值与GAT测量值相当,但测量值均比GAT测量值偏低。对于无眼表疾病患者,TX-20测量眼压的可重复性好。
英文摘要:
      Objective: To evaluate the accuracy and repeatability of measurement of intraocular pressure(IOP) with Canon TX-20 non-contact tonometer(TX-20) and Goldmann applanation tonometer(GAT). Methods: 120 eyes of 60 patients were enrolled in this study. For each eye, three consecutive IOP measurements were performed with TX-20 and GAT by the same examiner in 5minutes. The subjects were divided into three groups according to their IOP measured by GAT: group A(low group)(IOP<11mmHg,1kpa=7.5mmHg),group B(normal group)(11≤IOP≤21), group C(high group) (IOP>21mmHg). The accuracy of TX-20 was compared with GAT. The repeatability of TX-20 was performed by comparing three consecutive measurements. Results: Compared with the measured values of GAT, the 3 groups of A, B, C measured by TX-20 were lower. 95% confidence interval (95%CI) of the difference between them were 0.31 ~ 2.95mmHg, 1.09 ~ 4.38mmHg, 0.28 ~ 7.28mmHg, respectively. The repeatability of measurement of TX-20 was good, because of the difference between the 3 consecutive measurements were not statistically significant (P > 0.05). Conclusions: The TX-20 tonometer provided similar results to the Goldmann for low and normal group,but smaller than GAT. The repeatability of measurement of TX-20 was good.
查看全文  查看/发表评论  下载PDF阅读器
关闭